11.12 Adjustment mode
In weighing mode press the MENU button and keep it pressed until the acoustic
signal gets mute. The setup menu is displayed.
Use the navigation buttons
to select the menu item „Calibration mode“.
Acknowledge using PRINT button, the current setting is displayed.
Use the navigation buttons to select the desired setting.
! In models PLT-M only automatic adjustment is available.
Display
Description
Procedure see chap.
Internal
adjustm.
Adjustment with internal
weight after pressing the
CAL button
+
Chap. 7.1
External
adjustm.
Adjustment with external
adjustment weight
+
Chap. 7.3
Technical
adjustm.
Overwrite internal
adjustment weight
+
Chap. 7.4
Autom. adjustm.
Automatic adjustment with
internal adjustment weight
+
Chap. 7.2
+
or 8.1 (PLT-M)
Take over selection using the PRINT button.
Weighing balance returns to menu. Either make more settings in the menu or go
back to weighing mode as follows.
Press the MENU button and keep it pressed until the acoustic signal gets mute.
The balance returns automatically into weighing mode.