(9434) Enable Asynchronous Messages

Causes the unit to send the host the asynchronous messages when they are ready. By
default this option is disabled on startup. Once it is enabled, the messages are sent when
they are ready until the messages are disabled or until power is removed from the unit.

Battery Voltage

BD 01 (unsigned 16-bit in hundreds of volts - 4.45 V)

Checksum

59

Command String

03 02 10 94 34 01 01 00 01 20

Preamble

03 02 10

Command Code

94 34

Type code

01 (command being sent)

Data size

01 00 (1 byte being sent)

Data

01

Checksum

20

Response String

03 02 10 94 34 02 00 00 21

Preamble

03 02 10

Command Code

94 34

Type Code

02 (good response)

Data Size

00 00 (0 bytes returned)

Data

none

Checksum

21