beautypg.com

9435) set default thermometer mode, Command code (hex), Data – Welch Allyn SureTemp PlusModule - User Manual User Manual

Page 53: Response data, Notes

background image

SureTemp Plus Module OEM Implementation

Communications Protocol

Welch Allyn OEM Technologies

Confidential 53

(9435) Set Default Thermometer Mode

Causes the module to read the default thermometer mode from the host. The default mode
specifies whether the device enters normal predictive mode or monitor mode when the probe
is pulled.

Command Code (hex)

9435

Data

Size (bytes)

Type

Description

1

int8

0: Normal (Predictive)
1: Monitor

Response Data

Size (bytes)

Type

Description

0

N/A

None

Notes

Use the “Save Calibration Data to EEPROM” command (page 41) to save permanently to
EEPROM.