beautypg.com

9416) send 690 event log, Command code (hex), Data – Welch Allyn SureTemp PlusModule - User Manual User Manual

Page 49: Response data, 941a) send programmable ambient bias

background image

SureTemp Plus Module OEM Implementation

Communications Protocol

Welch Allyn OEM Technologies

Confidential 49

(9416) Send 690 Event Log

Causes the module to send between 0 and 10 bytes of data back to the host depending on
the number of errors in the event log. 10 is the maximum size of the event log, but if it is
partially empty then the response could be less then 10 bytes in length.

Command Code (hex)

9416

Data

Size (bytes)

Type

Description

0

N/A

None

Response Data

Size (bytes)

Type

Description

10

uint8

All errors are 1-byte. See page 71 for a mapping
between log numbers and display number.

(941A) Send Programmable Ambient Bias

Causes the module to send the programmable ambient bias to the host.

Command Code (hex)

941A

Data

Size (bytes)

Type

Description

0

N/A

None

Response Data

Size (bytes)

Type

Description

4

float32

The device’s ambient bias. It is different in the OEM