Packet Structure

SureTemp Plus Module OEM Implementation

36

Confidential

Welch Allyn OEM Technologies

TEMP (19 data bytes)

Type

Description

Values/Units

float32

“Final/Predicted” temperature

in

°

Kelvin

float32

Current Probe Temperature

in

°

Kelvin

uint32

Sample Counter - 173 mS/sample

int8

Method

0: Normal
1: Monitor
2: Clinical

int8

Algorithm

0: Oral
1: Rectal
2: Pediatric Axillary
3: Cal Key
4: Adult Axillary

int8

SureTemp state

0: Idle
1: Not Ready
2: Ready
3: Predict
4: Predict display
5: Predict position
6: Final
7: Final position
8: Final over temp
9: Final over temp position
10: Final under temp
11: Final under temp position
12: No predict (monitor mode)
13: End normal status
20: Error status

int16

Error code

0: No error
See Error Codes (page 65)

boolean

active

Probe out and Active

int8

Physical probe type connected

0: Oral/Axillary
1: Cal Key
2: Rectal
3: No Probe