Commands and Responses

SureTemp Plus Module OEM Implementation

48

Confidential

Welch Allyn OEM Technologies

(9413) Send Mode Pause For Clinical Device

Causes the module to send the number of milliseconds between the final predict temperature
and its transfer into monitor mode to the host. The host can control the transfer time from
predict mode to monitor mode.

Command Code (hex)

9413

Data

Size (bytes)

Type

Description

0

N/A

None

Response Data

Size (bytes)

Type

Description

2

int16

0 = Disabled; do not automatically switch to monitor
mode after a predict. This is because without a
pause the user would never see the final predicted
temperature.
Limits 0-10,000 where 10,000 = 10 seconds max.
(Value in milliseconds).

(9415) Send Default Thermometer Mode

Causes the module to send to the host the default thermometer mode. This allows the user to
specify which method of taking temperature readings we enter when the probe is pulled,
normal predictive mode, or monitor mode.

Command Code (hex)

9415

Data

Size (bytes)

Type

Description

0

N/A

None

Response Data

Size (bytes)

Type

Description

1

int8

0: Normal (Predictive)
1: Monitor