(9433) Set Mode Pause For Clinical Device

Enables the host to set the number of milliseconds the device pauses after a predict
temperature before switching to monitor mode. The transfer to monitor mode occurs after the
host-specified delay.
This is appropriate during clinical testing when comparing a predicted temperature to a
monitored temperature. It makes the final predict measurement available for a period before
the module switches to monitored mode and begins sending new temperature values.

Command Code (hex)

9433

Data

Size (bytes)

Type

Description

2

int16

0 = Disabled
Do not automatically switch to monitor mode after a
predict; without a pause the user would never see
the final predicted temperature.

Response Data

Size (bytes)

Type

Description

0

N/A

None

Notes

Use the “Save Calibration Data to EEPROM” command (page 41) to save permanently to