SureTemp Plus Module OEM Implementation

Communications Protocol

Welch Allyn OEM Technologies

Confidential 59

(2A12) Set Temperature Method

Causes the unit to read the current temperature method from the host. This specifies whether
the device enters into normal predictive or monitor mode when a probe is pulled.
Example:

Command String

03 02 10 2A 12 01 01 00 01 AC

Preamble

03 02 10

Command Code

2A 12

Type code

01 (command being sent)

Data size

01 00 (1 byte being sent)

Data

01 (1 byte where 01 = Monitor, 00 = Normal predict)

Checksum

AC

Response String

03 02 10 2A 12 02 00 00 AD

Preamble

03 02 10

Command Code

94 34

Type Code

02 (good response)

Data Size

00 00 (0 bytes returned)

Data

none

Checksum

AD