beautypg.com

Commands and responses, 2900) send serial number, Command code (hex) – Welch Allyn SureTemp PlusModule - User Manual User Manual

Page 38: Data, Response data, 2913) send temperature method

background image

Commands and Responses

SureTemp Plus Module OEM Implementation

38

Confidential

Welch Allyn OEM Technologies

Commands and Responses

(2900) Send Serial Number

Causes the module to send its serial number to the host.

Command Code (hex)

2900

Data

Size (bytes)

Type

Description

0

N/A

None

Response Data

Size (bytes)

Type

Description

11

ansichar

Module Serial Number
(10 ASCII characters + null terminator)

(2913) Send Temperature Method

Causes the module to send to the host the current temperature method. This flag is used to
specify whether the current temperature-taking method is normal predictive or monitor.

Command Code (hex)

2913

Data

Size (bytes)

Type

Description

0

N/A

None

Response Data

Size (bytes)

Type

Description

1

int8

0: Normal (predictive)
1: Monitor