Commands and Responses

SureTemp Plus Module OEM Implementation

52

Confidential

Welch Allyn OEM Technologies

(9434) Set Asynchronous Messages Enabled

Causes the module to send asynchronous messages to the host when they are ready.
By default this option is disabled on startup of the system. Once it is enabled, the temp record
is sent every 173 milliseconds until one of the following occurs:

the temperature cycle becomes inactive

a command is sent to disable the mode

power is removed from the module

In asynchronous mode, the module might not respond immediately to messages from the
host. Avoid continuous polling of the module when asynchronous messages are enabled and
a temperature cycle is active.

Command Code (hex)

9434

Data

Size (bytes)

Type

Description

1

int8

0: To disable asynchronous messages
1: To enable 33-byte asynchronous messages
2: To enable 19-byte asynchronous messages

Response Data

Size (bytes)

Type

Description

0

N/A

None

Notes

Once enabled, asynchronous messages are sent when they are ready. (The TEMP record is
the only asynchronous message; see page 35 and page 36.)
Data description (0, 1, or 2) can be switched at any time.