Guidelines
SureTemp Plus Module OEM Implementation
22
Confidential
Welch Allyn OEM Technologies
Figure 8. Acceptable Ways to Display the Rectal Mode
Monitor Mode
Monitor Mode continuously displays the temperature of the probe for as long as the probe
remains in place at the measurement site and remains within the operating patient temperature
range. The patient’s oral and rectal temperature reaches final equilibrium in approximately three
minutes in the Monitor Mode. Axillary temperature reaches equilibrium in approximately five
minutes in the Monitor Mode. Oral/Axillary and Rectal measurement modes time out after10
minutes of Monitor mode.
The device enters Monitor Mode for several reasons:
•
If the probe is withdrawn from the probe well and is not replaced within 60 seconds of
inactivity.
•
When a patient’s temperature in Normal Mode cannot correctly be measured.
•
Directly from a host command (for the purpose of testing, etc).
Monitor Mode Indication
Figure 9. Acceptable Ways to Display the Monitor Mode
Preferred
To indicate the monitor mode, display a snail icon (Figure 9) and the
word MONITOR.
Acceptable
If a graphical display cannot be accomplished, display the word
MONITOR or the abbreviation M.
Not acceptable
To provide no indication, or to use any other abbreviation or
inappropriate indicator, is not acceptable.
RECTAL
REC
MONITOR
M