Commands and Responses

SureTemp Plus Module OEM Implementation

44

Confidential

Welch Allyn OEM Technologies

(2A37) Set Default Thermometer Probe Mode

Causes the module to set the default probe mode from the host. This flag is used to define
what algorithm the device defaults to when an oral/axillary probe is connected to the device.

Command Code (hex)

2A37

Data

Size (bytes)

Type

Description

1

int8

0: Oral/Axillary
1: Pediatric Axillary
2: Adult Axillary
3: Use Last Algorithm

Response Data

Size (bytes)

Type

Description

0

N/A

None

Notes

Use the “Save Calibration Data to EEPROM” command defined in for (2A06) Save Data to
EEPROM on page 53 to save permanently to EEPROM.

(3000) Restart Module

Causes the module to reboot.

Command Code (hex)

3000

Data

Size (bytes)

Type

Description

0

N/A

None

Response Data

Size (bytes)

Type

Description

0

N/A

None