beautypg.com

3604) set temperature probe, Command code (hex), Data – Welch Allyn SureTemp PlusModule - User Manual User Manual

Page 47: Response data, Notes, 9401) send thermometer status no clear

background image

SureTemp Plus Module OEM Implementation

Communications Protocol

Welch Allyn OEM Technologies

Confidential 47

(3604) Set Temperature Probe

Causes the unit to simulate the removal or replacement of the temperature probe.

Command Code (hex)

3604

Data

Size (bytes)

Type

Description

1

boolean

0: Probe in holder (not ready for reading)
1: Probe removed from holder (ready for reading)

Response Data

Size (bytes)

Type

Description

0

N/A

None

Notes

To use this command, the temperature probe should normally be locked.

(9401) Send Thermometer Status No Clear

Causes the module to return a status block. The status block indicates what the module is
doing and reports any error conditions.

Command Code (hex)

9401

Data

Size (bytes)

Type

Description

0

N/A

None

Response Data

Size (bytes)

Type

Description

4

TEMP_STATUS Status

block

see TEMP_STATUS (page 34) for details

Notes

This command does not clear POST errors.