beautypg.com

9401) send thermometer status no clear – Welch Allyn SureTemp PlusModule - User Manual User Manual

Page 61

background image

SureTemp Plus Module OEM Implementation

Communications Protocol

Welch Allyn OEM Technologies

Confidential 61

(9401) Send Thermometer Status No Clear

Causes the unit to return a status block. The status block indicates what the unit is doing and
reports any error conditions.
Example:

The data size of 04 00 is a word consisting of a value of 0004. The status of FF 00 00 EC is
two words. (See TEMP_STATUS structure, page 34.)

The “POST Status" of FF 00 is the value of 00FF, which is 255 POST not yet available.

The "POST Tests" field of 00 EC is the value of EC00 (L H format) where the upper bit of
the E represents bit 15. (EC00 is 1110 1100 0000 0000.)

The bit mapping for the tests: 0: not performed, 1: performed

15: Data Bus Test - 1
14: Address Bus Test -1
13: Device Physical Memory Address Test- 1
12: Segments Test- 0
11: Heater Analog Test- 1
10: Heater Fail-safe Test- 1
09: Spare- 0
08: Spare- 0
07: Spare- 0
06: Spare- 0

Command String

03 02 10 94 01 01 00 00 55

Preamble

03 02 10

Command Code

94 01

Type code

01 (command being sent)

Data size

00 00 (0 bytes being sent)

Data

none

Checksum

55

Response String

03 02 10 94 01 02 04 00 FF 00 00 EC 65

Preamble

03 02 10

Command Code

94 01

Type Code

02 (good response)

Data Size

04 00 (4 bytes returned)

Data

FF 00 00 EC = a two-word status (See below.)

Checksum

65