9401) send thermometer status no clear – Welch Allyn SureTemp PlusModule - User Manual User Manual
Page 61

SureTemp Plus Module OEM Implementation
Communications Protocol
Welch Allyn OEM Technologies
Confidential 61
(9401) Send Thermometer Status No Clear
Causes the unit to return a status block. The status block indicates what the unit is doing and
reports any error conditions.
Example:
The data size of 04 00 is a word consisting of a value of 0004. The status of FF 00 00 EC is
two words. (See TEMP_STATUS structure, page 34.)
The “POST Status" of FF 00 is the value of 00FF, which is 255 POST not yet available.
The "POST Tests" field of 00 EC is the value of EC00 (L H format) where the upper bit of
the E represents bit 15. (EC00 is 1110 1100 0000 0000.)
The bit mapping for the tests: 0: not performed, 1: performed
15: Data Bus Test - 1
14: Address Bus Test -1
13: Device Physical Memory Address Test- 1
12: Segments Test- 0
11: Heater Analog Test- 1
10: Heater Fail-safe Test- 1
09: Spare- 0
08: Spare- 0
07: Spare- 0
06: Spare- 0
Command String
03 02 10 94 01 01 00 00 55
Preamble
03 02 10
Command Code
94 01
Type code
01 (command being sent)
Data size
00 00 (0 bytes being sent)
Data
none
Checksum
55
Response String
03 02 10 94 01 02 04 00 FF 00 00 EC 65
Preamble
03 02 10
Command Code
94 01
Type Code
02 (good response)
Data Size
04 00 (4 bytes returned)
Data
FF 00 00 EC = a two-word status (See below.)
Checksum
65