beautypg.com

3101) send digital input status, Command code (hex), Data – Welch Allyn SureTemp PlusModule - User Manual User Manual

Page 45: Response data

background image

SureTemp Plus Module OEM Implementation

Communications Protocol

Welch Allyn OEM Technologies

Confidential 45

(3101) Send Digital Input Status

Causes the module to send the status of all the device’s digital inputs to the host. This is a
short representation of all of the switches and buttons on the device. 1 is pressed or closed,
and 0 is released, or open.

Command Code (hex)

3101

Data

Size (bytes)

Type

Description

0

N/A

None

Response Data

Size (bytes)

Type

Description

2

uint16

Each bit represents a different button or switch on

the device.

15: Reserved
14: Reserved
13: Reserved
12: Reserved
11: Reserved
10: Probe switch
09: Probe attached
08: Probe well inserted
07: Spare
06: Spare
05: Spare
04: Spare
03: Spare
02: Spare
01: Spare
00: Spare