941c) send temperature record - 33 data bytes – Welch Allyn SureTemp PlusModule - User Manual User Manual
Page 64

Command and Response Strings: Examples
SureTemp Plus Module OEM Implementation
64
Confidential
Welch Allyn OEM Technologies
(941C) Send Temperature Record - 33 Data Bytes
Causes the unit to send to the host the current temperature record. The temperature record is
updated every 173 milliseconds.
Command String
03 02 10 94 1C 01 00 00 3A
Preamble
03 02 10
Command Code
94 1C
Type code
01 (command being sent)
Data size
00 00 (0 bytes being sent)
Data
none
Checksum
3A
Response String
03 02 10 94 1C 02 21 00 00 00 00 00 76 AF 99 43 4E 00 23 00 00 00
02 00 00 01 00 EF 9E 52 46 5d 4D 94 43 51 12 3D 46 43 BD 01 59
Preamble
03 02 10
Command Code
94 1C
Type Code
02 (good response)
Data Size
21 00 (33 Decimal Bytes)
Data
Final/Predicted Temperature
00 00 00 00 (IEEE 32-bit Float = 0.0
°
K)
Current Probe Temp
76 AF 99 43 (IEEE 32-bit Float = 307.370789
°
K)
Current Heater Count/PWM
4E 00 (Not important for the OEM user)
Sample Counter
23 00 (unsigned 16-bit = 35 Decimal)
Method
00 (8-bit where 0 is Predict Method/Mode)
Algorithm
00 (8-bit where 0 is Oral)
SureTemp state
02 (8-bit where 2 is Ready)
Error code
00 00 (unsigned 16-bit where 0 means no error)
Active
01 (Boolean where 1 = Active)
Physical probe type connected
00 (8-bit where 0 is Oral/Axillary Probe Type)
Probe Resistance
EF 9E 52 46 (Not important for the OEM user)
Ambient Temp
5D 4D 94 43 (IEEE 32-bit Float = 296.604401
°
K)
PTB Resistance
51 12 3D 46 43 (Not important for the OEM user)