Drucker Diagnostics Paralens Advance User Manual

Page 61

background image

Table A-2-1

WITHIN-RUN PRECISION, QBC AUTOREAD PLUS SYSTEM

9

QBC VENOUS BLOOD TUBES

N = 10

per sample

HCT

HB

MCHC

PLT

WBC

#GRANS

%GRANS

#L/M

%L/M

SAMPLE 1

MEAN

47.6

15.9

33.5

155

7.1

4.6

65

2.5

35

% CV

1.87

0.74

2.17

8.91

3.16

2.71

2.02

5.86

3.78

SAMPLE 2

MEAN

23.1

7.4

31.8

288

6.7

4.9

73

1.8

27

% CV

1.44

0.66

1.22

3.27

2.33

3.07

1.41

3.69

3.85

SAMPLE 3

MEAN

31.9

10.4

32.5

550

11.2

7.9

71

3.3

29

% CV

1.5

1.27

1.29

3.59

5.41

7.3

2.86

6.49

6.96

SAMPLE 4

MEAN

52.3

16.3

31.1

745

16.1

10.6

66

5.6

34

% CV

1.4

1.35

0.96

4.68

6.42

5.86

1.43

8.09

2.77

SAMPLE 5

MEAN

41.6

13.6

32.8

91

5.3

3.6

68

1.7

32

% CV

1.73

1.45

0.8

5.34

6.72

6.2

1.85

9.02

4

QBC CAPILLARY BLOOD TUBES

N = 10

per sample

HCT

HB

MCHC

PLT

WBC

#GRANS

%GRANS

#L/M

%L/M

SAMPLE 1

MEAN

49.1

15.3

31.2

144

8.4

5.9

70

2.5

30

% CV

1.1

0.9

0.9

13.1

6.7

10.1

4.3

7.4

10.2

SAMPLE 2

MEAN

31.9

9.7

30.3

522

11.6

8.1

70

3.5

30

% CV

2.0

1.1

1.7

3.7

4.5

4.1

1.1

6.1

2.6

SAMPLE 3

MEAN

54.9

16.4

29.9

582

15.6

10.3

66

5.4

34

% CV

0.9

1.2

1.0

6.6

2.8

2.9

2.7

7.0

5.2

SAMPLE 4

MEAN

42.8

12.9

30.2

88

5.7

3.8

66

2

34

% CV

0.8

0.7

0.6

9.9

5.6

5.0

2.7

9.4

5.3

SAMPLE 5

MEAN

40.8

12.5

30.7

197

5.9

3.9

67

2

33

% CV

1.9

1.5

1.1

8.3

7.5

9.0

1.7

4.2

3.5

A-2.4 

 

SPECIFIC PERFORMANCE

CHARACTERISTICS

A-2.4.1 Precision

Data on within-run reproducibility of the QBC
Autoread Plus System from five blood specimens with
QBC venous tubes and five blood specimens with
QBC capillary tubes are shown in Table A-2-1. The

means and % C.V.’s below are based on replicate
tests of 10 tubes per sample; e.g., for Sample 1,
n=10, etc.

A-2-3