Drucker Diagnostics Paralens Advance User Manual
Page 41
![background image](/manuals/601320/41/background.png)
Section 6
Test Procedure – QBC AccuTube
TM
6.1 MATERIALS PROVIDED
The Autoread Plus Hematology System consists
of the following instruments and accessories with
which to perform hematology tests:
• QBC Autoread Plus Analyzer with software
cartridge
• QBC Power Pack
• QBC Capillary Centrifuge
• Printer
• Calibration Check Rod
• Workstation
• Forceps
• QBC Pipetter with AccuTube Spacer
6.2 MATERIALS REQUIRED BUT NOT PROVIDED
•QBC AccuTubes
100 Tests
423406
• Lint-free tissue
6.3
HEMATOLOGY TEST PROCEDURE WITH
QBC ACCUTUBE
• Be sure QBC blood tubes have not exceeded
their labeled expiration date or open vial
stability.
• Maintain laboratory temperature at 68° to 98°F
(20° to 37°C).
• Venous blood: specimen must be well-mixed
and at room temperature.
• Capillary blood: collect only from free-flowing
finger puncture or plantar surface of heel in
infants less than one year old.
WARNING
Acridine orange reagent may be toxic; do not
ingest. Avoid contact with skin, eyes, and
clothing.
WARNING
Blood specimens may contain the Hepatitis B
Virus (HBV), Hepatitis C Virus (HCV), Human
Immuno deficiency Virus (HIV), or other disease-
causing agents. Handle all patient specimens
as potential biohazards capable of transmitting
infection. Wear appropriate personal protective
equipment, including gloves, when collecting
and processing blood.
WARNING
QBC blood tubes are made of glass. Be careful
when handling and preparing tubes to prevent
breakage and possible injury. Inspect QBC
AccuTubes before use. Do not use cracked or
scratched tubes.
6-1