Drucker Diagnostics Paralens Advance User Manual
Page 10

2.3 SETUP PROCEDURES (continued)
2.3.4 Power Requirements
Plug the power cords of the power pack and
centrifuge into the grounded electrical receptacles
rated for the line voltage and frequency specified
on their respective data plates. For centrifuge
voltage tolerances, consult the operator’s manual
of the QBC Capillary Centrifuge supplied with the
System.
For proper operation of the Autoread Plus analyzer,
the power source must deliver voltages within the
limits specified below.
Model No.
Voltage Tolerance
424590
Autoread Plus
90-265 VAC
Power Pack
50-60 Hz
If the line voltage is known to fluctuate outside
the above tolerances, notify the installer; a special
voltage regulator may be required.
The analyzer is electrically energized by the
rocker-type POWER switch in the back of the unit
(see Figure 2-3). The switch is labeled with two
positions: [0] = Off and [1] = On. When power is on,
a green LED Power indicator on the front display
panel is illuminated.
CAUTION
Connect the power cord only to a 3-wire
grounded receptacle delivering the voltage and
frequency specified on the data plate of the
power pack. Where only a 2-wire receptacle
is available, have it replaced by a qualified
individual and in accordance with all specified
electrical codes. If an extension cord is
required, use only a 3-wire grounded cord with
the proper voltage rating.
2-3