4.3  TEST ALERTS

4.3.1 Flashing Values and Dashes

When a test value flashes on and off on the display
panel, an asterisk (*) will appear on the printout after
the numerical value. Flashing values indicate that
the result is outside the validated test range of the
parameter. (See Table 4-1).

Dashes on the panel and on the printout mean
that the computed test value is outside the display
range of the instrument, or that a packed cell layer
is too small to measure. In such cases, dashes and
no test value and bar graph point will appear.

Table 4-1 shows out-of-range points for QBC
AccuTubes. With QBC AccuTubes, elevated cell
counts are indicated by the statement OUT OF
RANGE on the printout and by the following
displays:

• High WBC – Display flashes 99.9 (instead of ---)

when total WBC count is over 99.9 × 10

9

/L.

• High Granulocyte Count – Display flashes 99.9

when Granulocyte count is over 99.9 × 10

9

/L.

• High Lymph/Mono Count – Display flashes 99.9

when Lymph/Mono count is over 99.9 × 10

9

/L.

• High Platelet Count – Display flashes 999 when

Platelet count is over 999 × 10

9

/L.

See Appendix A-2 for out-of-range points for other
QBC tube types.

4.3.2 Special HB-MCHC Conditions

No HB and MCHC values will appear if any of the
conditions listed in Table 4-2 occur. In such cases,
the examination of a peripheral blood smear is
recommended to determine the nature of the red
cell abnormality.

4-6

Table 4-1.

Out-of-Range Points for

Flashing Test Values – AccuTube

HCT

< 15% or > 65%

 PLT*

 < 20 or > 999

 WBC*

 < 1.6 or > 99.9

 GRANS*

< 0.8 or > 70

.

 LYMPH/MONOS*

 < 0.8 or > 99.9

*10

9

/L

Table 4-2.

HB-MCHC No-Value Conditions

HCT

≤ 15% or ≥ 65%

HB

≤ 5 g/dL or ≥ 20 g/dL

MCHC

≤ 25 g/dL or ≥ 37.3 g/dL