Drucker Diagnostics Paralens Advance User Manual
Page 29
![background image](/manuals/601320/29/background.png)
4.3 TEST ALERTS
4.3.1 Flashing Values and Dashes
When a test value flashes on and off on the display
panel, an asterisk (*) will appear on the printout after
the numerical value. Flashing values indicate that
the result is outside the validated test range of the
parameter. (See Table 4-1).
Dashes on the panel and on the printout mean
that the computed test value is outside the display
range of the instrument, or that a packed cell layer
is too small to measure. In such cases, dashes and
no test value and bar graph point will appear.
Table 4-1 shows out-of-range points for QBC
AccuTubes. With QBC AccuTubes, elevated cell
counts are indicated by the statement OUT OF
RANGE on the printout and by the following
displays:
• High WBC – Display flashes 99.9 (instead of ---)
when total WBC count is over 99.9 × 10
9
/L.
• High Granulocyte Count – Display flashes 99.9
when Granulocyte count is over 99.9 × 10
9
/L.
• High Lymph/Mono Count – Display flashes 99.9
when Lymph/Mono count is over 99.9 × 10
9
/L.
• High Platelet Count – Display flashes 999 when
Platelet count is over 999 × 10
9
/L.
See Appendix A-2 for out-of-range points for other
QBC tube types.
4.3.2 Special HB-MCHC Conditions
No HB and MCHC values will appear if any of the
conditions listed in Table 4-2 occur. In such cases,
the examination of a peripheral blood smear is
recommended to determine the nature of the red
cell abnormality.
4-6
Table 4-1.
Out-of-Range Points for
Flashing Test Values – AccuTube
HCT
< 15% or > 65%
PLT*
< 20 or > 999
WBC*
< 1.6 or > 99.9
GRANS*
< 0.8 or > 70
.
LYMPH/MONOS*
< 0.8 or > 99.9
*10
9
/L
Table 4-2.
HB-MCHC No-Value Conditions
HCT
≤ 15% or ≥ 65%
HB
≤ 5 g/dL or ≥ 20 g/dL
MCHC
≤ 25 g/dL or ≥ 37.3 g/dL