Step 2: Roll Tube Between Fingers to Mix

Gently roll tube between fingers at least 10 times

or for at least 5 seconds, keeping unsealed end

slightly above horizontal (D).

Proceed promptly to

Step 3.

Step 3: Insert Float

With 424240 Tray – Slide unsealed end of tube over

tip of pre-positioned float (E1) and push until float is

inside tube as far as possible. Gently lift closure end

of tube until float releases from its tray slot. Raise

unsealed end of tube slightly above horizontal to

prevent float from falling out. If necessary, press

end of float against clean surface until float is inside

tube.

NOTE: Never touch floats with fingers or

gloves. Use forceps to handle loose or dropped

floats.

With 424237 Box – Using forceps supplied with

analyzer, pick up float from well. Insert forceps-held

float into unsealed end of tube (E2). With forceps,

tap float into tube.

Never touch float with fingers

or gloves.

Note: After inserting float, tubes can be held for

centrifuging provided they are stored vertically,

closure down in workstation. A maximum of

15 minutes is allowable between insertion of float

and centrifugation.

Step 4: Centrifuge for 5 Minutes

Place blood tubes in centrifuge. Spin down

according to instructions in centrifuge manual.

When centrifugation is complete,

promptly remove

tubes.

Time Delay Between Completion of

Centrifugation and Tube Reading: Centrifuged

QBC tubes are stable for up to 4 hours prior

to reading, if stored vertically (closure down) in

workstation, away from heat and intense light.

IMPORTANT:

• Excessive heat may disturb cell interfaces in

centrifuged blood tube. Avoid picking up or

handling spun tube below plasma column.

Do not place centrifuged tubes on warm

surfaces or under intense light.

• Do not leave blood tubes on rotor after

centrifugation. Remove and read

immediately, or temporarily store in vertical,

closure-down position.

• Do not twist or move tube closures after

centrifugation.