Drucker Diagnostics Paralens Advance User Manual

Page 34

background image

4-11

ERROR MESSAGE

CAUSE AND ACTION

Lamp Test

Failed

Cause: Display message will flash if software determines the Lamp is either open or shorted.
Action: Request service.

Checksum

Error 1

Cause: Defective cartridge.
Action: Turn Power switch off, remove cartridge, and re-install, turn Power on and try again.
If error persists, order a replacement cartridge.

Checksum

Error 2

Cause: Defective cartridge.
Action: Turn Power switch off, remove cartridge, and re-install, turn Power on and try again.
If error persists, order a replacement cartridge.

Calibration Error

(backlash)

Cause: Loose motor coupling, or carriage not moving freely.
Action: Request service.

Filter Wheel

Error (1)

Also #2, #3, #4, #5 & #6

Cause: Defective filter wheel, defective LED, carriage not moving, or defective circuit board.
Action: Request service.

Position Error

[Remove Tube]

*With Buzzing noise

• • •

If message occurs on powering up, error is a holdover from the last procedure.
Cause: During carriage return, tube ejected improperly by collet.
Action: Open door; carefully remove tube with forceps. Follow display messages by pressing
[NEXT] key, then simultaneously pressing [5] and [6] keys.

Carriage Error

(no sensor)

• • •

Cause: Transport mechanism is not moving due to a: 1) defective power pack; 2) jammed
carriage; or 3) defective circuit board.
Action: Request service.

Rotation Error

Cause: Incorrect or damaged calibration check rod. Damaged collet. Optics require service.
Action: Verify use of correct Call Check Rod. Wipe clean and retest.
If message persists, request service.

Cal Rod Error (03)

Cause: Incorrect or damaged calibration check rod. Damaged collet. Optics require service.
Action: Verify use of correct Call Check Rod. Wipe clean and retest.
If message persists, request service.

Improper

QBC Tube Type

Cause: Mode on instrument incorrect for sample being tested (i.e. calibration check rod
tested in CBC or Insert QBC Tube mode).
Action: Remove tube and insert correct type... or access correct test mode.

*Cannot ID

QBC Tube Type

Cause: Severely hemolyzed specimen, severe platelet clumping on top of float, improperly
prepared tube, missing float, tube filled backwards, smudged tube, missing fill lines, or wrong
tube type.
Action: Remove tube and inspect. Retest or prepare and test new tube.

Error Locating

Bottom of RBC’s

Cause: Closure or stopper in QBC tube is incorrectly sealed, or LED is defective.
Action: Prepare and test new tube. If message persists, request service.

*Error

Locating Float

Cause: Severely hemolyzed specimen, defective float, float lodged in wrong part of tube,
improperly prepared tube, defective lamp, or defective LED.
Action: Remove tube and inspect. Retest or prepare and test new tube. If message persists,
request service.

DURING STARTUP AND SELF-TEST

DURING CALIBRATION CHECK

DURING HEMATOLOGY TESTS*

*Patient, Proficiency and Control Samples

Table 4-3

Error Messages and Codes

*For trouble shooting purposes refer to the print a diagnostic scan for technical services to interpret.