beautypg.com

LaMotte SMART3 Colorimeter User Manual

Page 54

background image

ENTERING A MULTIPLE POINT CALIBRATION

The SMART3 can directly create a 2 point calibration curve. (See Entering a

Two Point Calibration on page 47.) To create a multiple point calibration curve,

constants obtained from a linear regression of multiple data points can be

entered into the SMART 3.
1. Scan reactions of multiple concentrations at the appropriate wavelength in the

absorbance mode on the SMART3.
2. Plot the concentration (y axis) versus absorbance (x axis) in a program

capable of linear regression such as Excel.
3. Enter the constants obtained from the linear regression equation into the

SMART3.
For Example:

y = 0.001x

3

– 0.017x

2

+ 0.181x – 0.049

K0 = – 0.049

K1 = 0.181

K2 = – 0.017

K3 = 0.001

OR (Over Range) = 10

1. From the User Test menu,

press

or

to scroll

to

Enter Constants.

111 H2O

Name the Test
Select Vial/WL
STD Calibration

Enter Constants

12:00:00 001/500

2. Press

ENTER

to select

Enter

Constants.

K0=0.00000
K1=0.00000
K2=0.00000
K3=0.00000
OR=100.00000

12:00:00 001/500

52

SMART3 Colorimeter 11.10

Editing/Set Up