PROCEDURE

1. Press and hold

until colorimeter turns on.

2. Press

ENTER

to select TESTING MENU.

3. Select

ALL TESTS (or another sequence containing 085 Silica LR) from

TESTING MENU.

4. Scroll to and select 085 Silica LR from menu.
5. Rinse a clean tube (0290) with sample water. Fill to the 10 mL line with sample.
6. Insert tube into chamber, close lid and select SCAN BLANK. (See Note)
7. Remove tube from colorimeter. Add 6 drops *Silica Reagent #1 (V-4466). Cap

and invert to mix.

8. Add 12 drops of *Silica Reagent #2 (V-4467). Cap and mix. Wait 5 minutes.
9. Add 8 drops of *Silica Reagent #3 (V-4468). Cap and mix. Wait 2 minutes.
10. Use the 0.1 g spoon (0699) to add one measure of *Silica Reagent #4 (V-

6284). Cap and mix gently until powder has dissolved. Wait 5 minutes for full

color development.

11. At end of 5 minute waiting period, mix, insert tube into chamber, close lid and

select SCAN SAMPLE. Record result.

12. Press

to turn colorimeter off or press

EXIT

to exit to a previous menu or

make another menu selection.

NOTE: For best possible results, a reagent blank should be determined to account

for any contribution to the test result by the reagent system. To determine the

reagent blank, follow the above test procedure to scan a distilled or deionized

water blank. Then follow the above procedure to perform the test on a distilled or

deionized water sample. This test result is the reagent blank. Subtract the reagent

blank from all subsequent test results of unknown samples. It is necessary to

determine the reagent blank only when a new lot number of reagents are obtained.

SMART3 Test Procedures 11.10

SILICA, Low Range