Copper – low range – LaMotte SMART3 Colorimeter User Manual
Page 153
COPPER – LOW RANGE
BICINCHONINIC ACID METHOD • CODE 3640-SC
QUANTITY
CONTENTS
CODE
50
*Copper Tablets
*T-3808-H
*WARNING: Reagents marked with an * are considered to be potential health
hazards. To view or print a Material Safety Data Sheet (MSDS) for these reagents
go to www.lamotte.com. To obtain a printed copy, contact LaMotte by e-mail,
phone or fax.
The copper content of drinking water generally falls below 0.03 parts per million,
but copper levels as high as 1.0 part per million will give water a bitter taste.
Waters testing as high as 1.0 part per million copper have probably been treated
with a copper compound, like those used in the control of algae, or have become
contaminated from untreated industrial wastes. The addition of copper sulfate to
lakes causes an increase in the copper content of the sediments. Acid waters and
those high in free carbon dioxide may cause the corrosion or “eating away” of
copper, brass and bronze pipes and fi ttings. This corrosion results in the addition
of copper into the water supply.
APPLICATION:
Drinking, surface, and saline waters; domestic and
industrial wastes.
RANGE:
0.00–3.50 ppm Copper
MDL:
0.04 ppm
METHOD:
Copper ions form a purple complex with bicinchoninic
acid around pH 6-7, in proportion to the concentration of
copper in the sample.
SAMPLE HANDLING
& PRESERVATION:
Copper has a tendency to be adsorbed to the surface of
the sample container. Samples should be analyzed as
soon as possible after collection. If storage is necessary,
0.5 mL of 20% HCl per 100 mL of sample will prevent
“plating out.” However, a correction must be made to
bring the reaction into the optimum pH range.
INTERFERENCES:
High concentrations of oxidizing agents, calcium, and
magnesium interfere. Silver can also interfere.
SMART3 Test Procedures 11.10
COPPER, BCA, Tablet