PROCEDURE

1. Press and hold

until colorimeter turns on.

2. Press

ENTER

to select TESTING MENU.

3. Select

ALL TESTS (or another sequence containing 021 Chromium) from

TESTING MENU.

4. Scroll to and select 021 Chromium from menu.
5. Rinse a clean tube (0290) with sample water. Fill to the 10 mL line with sample.
6. Insert tube into chamber, close lid and select SCAN BLANK.
7. Remove tube from colorimeter. Use the 0.1g spoon (0699) to add one measure

of *Chromium Reagent Powder (V-6276). Cap and shake until powder

dissolves. Wait 3 minutes for full color development.

8. During waiting period, fold a piece of fi lter paper (0465) in half then half again

to form a cone. Push corners together to open end, and insert into funnel

(0459).

9. At the end of 3 minute waiting period, fi lter sample into a clean tube. Mix. Insert

tube into chamber, close lid and select SCAN SAMPLE. Record result.

10. Press

to turn colorimeter off or press

EXIT

to exit to a previous menu or

make another menu selection.

NOTES: To convert result to ppm chromate (CrO

4

–2

) multiply by 2.23. To convert

result to ppm sodium chromate (Na

2

CrO

4

) multiply by 3.12.

Highly buffered waters may give poor results and require a more careful pH

adjustment. Before adding *Chromium Reagent Powder, adjust pH of sample to

pH 3–4.

CHROMIUM, Hexavalent

SMART3 Test Procedures 11.10

Test P