PROCEDURE

1. Press and hold

until colorimeter turns on.

2. Press

ENTER

to select TESTING MENU.

3. Select

ALL TESTS (or another sequence containing 094 Surfactants) from

TESTING MENU.

4. Scroll to and select 094 Surfactants from menu.
5. Rinse a tube (0290) with sample water. Fill to 10 mL line with sample.
6. Insert the tube into chamber, close lid and select SCAN BLANK.
7. Remove the tube from colorimeter.
8. Use the 0.5 g spoon (0698) to add 0.5 g pH Adjustment Powder (4509). Cap

and mix until powder dissolves.

9. Use the 0.1 g spoon (0699) to add two measures of Sodium Chloride Reagent

(4877). Cap and mix until powder disintegrates.

10. Use the 1.0 mL pipet (0354) to add 2.0 mL of *DS Indicator (4508).
11. Cap and shake for 1 minute.

NOTE: Bubbles on the sides of the tube will interfere with the results. Swirl the

tube to remove bubbles if they are present.

12. Wait 5 minutes. DO NOT MIX.
13. Insert tube into chamber, close lid and select SCAN SAMPLE. Record result in

ppm LAS.

14. Press

to turn colorimeter off or press

EXIT

to exit to a previous menu or

make another menu selection.

NOTE: For best possible results, a reagent blank should be determined to account

for any contribution to the test result by the reagent system. To determine the

reagent blank, follow the above test procedure to scan a distilled or deionized water

blank. Then follow the above procedure to perform the test on a distilled or deionized

water sample. This test result is the reagent blank. Subtract the reagent blank from

all subsequent test results of unknown samples. It is necessary to determine the

reagent blank only when a new lot number of reagents is obtained.

SURFACTANTS

SMART3 Test Procedures 8.13