CALIBRATING LaMOTTE PRE-PROGRAMMED TESTS

The LaMotte Pre-Programmed Tests have been pre-calibrated. Recalibration of

the pre-programmed tests by the user is not possible. However, a procedure

to standardize the calibration can be performed to obtain the most accurate

readings or to meet regulatory requirements.
The LaMotte Pre-Programmed tests are standardized with one standard

solution. To standardize over the full range of the test, the concentration of

the standard should be chosen from the high end of the range. Alternatively, if

samples do not cover the full range of the test, a standard should be chosen

that is close to the concentration of the samples.
The standardization procedure should be followed as often as required by

regulations and laws for compliance monitoring.
In the example below, the Aluminum calibration will be standardized.

Prepare a standard solution to be tested. In this example, 0.30 ppm aluminum.

1. Press and briefl y hold

to turn the meter on. The

LaMotte logo screen will

appear for about 3 seconds

and the

Main Menu will

appear.

Main Menu

Testing Menu

Editing Menu
Run PC LINK

12:00:00 001/500

2. Press

ENTER

to select

Testing

Menu.

Testing Menu

All Test Menu

Sequence 1
Sequence 2
Sequence 3

12:00:00 001/500

3. Press

ENTER

to select

All Tests

Menu.

All Tests

001 Alkalinity UDV

002 Aluminum
003 Ammonia-N LRF
004 Ammonia-N LRS

12:00:00 001/500

SMART3 Colorimeter 11.10

29

Testing