PROCEDURE

1. Press and hold

until colorimeter turns on.

2. Press

ENTER

to select Testing Menu.

3. Select

ALL TESTS (or another sequence containing 048 H Peroxide Shock)

from TESTING MENU.

4. Scroll to and select 048 H Peroxide Shock from menu.
5. Use the pipet (0342) to add 5 drops of the sample water to a tube (0290).
6. Dilute to the 10 mL line with distilled or hydrogen peroxide-free water.
7. Insert the tube into chamber, close lid and select SCAN BLANK.
8. Remove the tube from colorimeter and add 4 drops of *Hydrogen Peroxide

Reagent #1 (6452). Cap and mix.

9. Add one *Hydrogen Peroxide LR Tablet (6454A). Crush tablet with Tablet

Crusher (0175). Cap and mix for 30 seconds. Solution will turn pink if hydrogen

peroxide is present. Wait 5 minutes for full color development.

10. At the end of 5 minute waiting period, mix, insert tube into chamber, close lid

and select SCAN SAMPLE. Record result.

11. Press

to turn the meter off or press

EXIT

to exit to a previous menu or

make another menu selection.

NOTES: For best possible results, a blank should be determined to account for

any contribution to the test result by the reagent system. To determine the reagent

blank, follow the above test procedure to scan a distilled or deionized water blank.

Then follow the above procedure to perform the test on a distilled or deionized

water sample. This test result is the reagent blank. Subtract the reagent blank from

all subsequent test results of unknown samples. It is necessary to determine the

reagent blank only when a new lot number of reagents is obtained.

For the most accurate results, the sample and reagents should be at 25±4°C.

HYDROGEN PEROXIDE, Shock