LaMotte SMART3 Colorimeter User Manual
Page 44
EDIT USER TESTS
If a test other than the LaMotte programmed tests is performed regularly, a
calibration for it may be entered in one of the 25 User Tests. These tests are
originally named “User Test 1 - 25”. It will be possible to rename the test, select
a wavelength, enter a new calibration, select the number of decimal places
used to display the results, and select the units. A User Test may be added for a
reagent system for which no precalibrated test exists. A calibration of a LaMotte
reagent system may also be entered. The calibration of a User Test can be
changed at any time.
The User Tests have the ability to handle 2 data points. The colorimeter will
determine the absorbance of the standards and calculate a response that
will be stored to determine the concentration of future samples of unknown
concentration. These standards should cover all the concentrations for the
range of the test being performed and be scanned beginning with the low
concentration and fi nishing with the high concentration (for more information
about this, see CALIBRATION CURVES, page 12). Prepare these standards prior
to entering a new calibration.
NOTE: A calibration procedure must be performed before using any of the User
Tests.
The User Tests can be placed in any of the alterable sequences using Edit
Sequences.
1. Press and briefl y hold
to turn the meter on. The
LaMotte logo screen will
appear for about 3 seconds
and the
Main Menu will
appear.
Main Menu
Testing Menu
Editing Menu
Run PC Link
12:00:00 001/500
2. Press
or
to scroll
to the
Editing Menu.
Main Menu
Testing Menu
Editing Menu
Run PC Link
12:00:00 001/500
42
SMART3 Colorimeter 11.10
Editing/Set Up