beautypg.com

Procedure – LaMotte SMART3 Colorimeter User Manual

Page 253

background image

PROCEDURE

1. Press and hold

until colorimeter turns on.

2. Press

ENTER

to select TESTING MENU.

3. Select

ALL TESTS (or another sequence containing 078 Phosphate LR) from

TESTING MENU.

4. Scroll to and select 078 Phosphate LR from menu.
5. Rinse a clean tube (0290) with sample water. Fill to the 10 mL line with sample.
6. Insert tube into chamber, close lid and select SCAN BLANK.
7. Remove tube from colorimeter. Use 1.0 mL pipet (0354) to add 1.0 mL of

*Phosphate Acid Reagent (V-6282). Cap and mix.

8. Use the 0.1 g spoon (0699) to add one measure of *Phosphate Reducing

Reagent (V-6283). Cap and mx until powder dissolves. Wait 5 minutes for full

color development. Solution will turn blue if phosphates are present.

9. At end of 5 minute waiting period, mix, insert tube into chamber, close lid and

select SCAN SAMPLE. Record result.

10. Press

to turn colorimeter off or press

EXIT

to exit to a previous menu or

make another menu selection.

SMART3 Test Procedures 11.10

PHOSPHATE, High Range

Test P

rocedures