PROCEDURE

1. Use Indicator, Range, & Test Name chart to select the indicator, corresponding

to anticipated pH range and to determine corresponding test name to select

from colorimeter menu.

2. Press and hold

until colorimeter turns on.

3. Press

ENTER

to select TESTING MENU.

4. Select

ALL TESTS (or another sequence containing the appropriate pH test

name) from TESTING MENU.

5. Scroll to and select the appropriate pH test name from menu.
6. Rinse a clean tube (0290) with sample water. Fill to the 10 mL line with sample.
7. Insert tube into chamber, close lid and select SCAN BLANK.
8. Remove tube from colorimeter. Use the 0.5 mL pipet (0369) to add exactly 0.5

mL of the pH indicator for the chosen range. Cap and mix.

9. Insert tube into chamber, close lid and select SCAN SAMPLE. Record result.
10. Press

to turn colorimeter off or press

EXIT

to exit to a previous menu or

make another menu selection.

SMART3 Test Procedures 11.10

pH