PROCEDURE - OZONE ONLY

1. Press and hold

until colorimeter turns on.

2. Press

ENTER

to select TESTING MENU.

3. Select

ALL TESTS (or another sequence containing 070 Ozone DPD) from

TESTING MENU.

4. Scroll to and select 070 Ozone DPD from menu.
5. Rinse a tube (0290) with sample water. Fill to 10 mL with sample.
6. Insert the tube into chamber, close lid and select SCAN BLANK.
7. Remove tube from colorimeter.
8. Add 5 drops *DPD #3 Total Chlorine Reagent (6743). Swirl to mix.
9. Add 5 drops DPD #1 A Free Chlorine Reagent (6740) and 5 drops *DPD #1B

Free Chlorine Reagent (6741).

10. Cap and invert to mix. Make reading within 30 seconds.
11. Insert tube into chamber, close lid and select SCAN SAMPLE. Record result

as mg/L ozone.

12. Press

to turn the meter off or press

EXIT

to exit to a previous menu or

make another menu selection.

NOTES: For best possible results, a blank should be determined to account for

any contribution to the test result by the reagent system. To determine the reagent

blank, follow the above test procedure to scan a distilled or deionized water blank.

Then follow the above procedure to perform the test on a distilled or deionized

water sample. This test result is the reagent blank. Subtract the reagent blank from

all subsequent test results of unknown samples. It is necessary to determine the

reagent blank only when a new lot number of reagents are obtained.

OZONE, DPD, Liquid

SMART3 Test Procedures 11.10