beautypg.com

Procedure – LaMotte SMART3 Colorimeter User Manual

Page 210

background image

PROCEDURE

1. Press and hold

until colorimeter turns on.

2. Press

ENTER

to select TESTING MENU.

3. Select

ALL TESTS (or another sequence containing 061 Molybdenum HR)

from TESTING MENU.

4. Scroll to and select 061 Molybdenum HR from menu.
5. Fill clean tube (0290) to 10 mL line with sample water.
6. Insert tube into chamber, close lid and select SCAN BLANK.
7. Remove tube from colorimeter. Use a 1.0 mL pipet (0354) to add 1.0 mL of

*Mo Buffer (3997). Cap and mix.

8. Use a second 1.0 mL pipet (0372) to add 1.0 mL of *Molybdenum Oxidizing

Reagent (6485). Cap and mix.

9. Use 0.05 g spoon (0696) to add one measure of Molybdenum Indicator

Powder (6486). Cap and mix until powder dissolves. Solution will turn yellow if

molybdenum is present. Mix the tbe to remove bubbles.

10. Insert tube into chamber, close lid and select SCAN SAMPLE. Record result.
11. Press

to turn colorimeter off or press

EXIT

to exit to a previous menu or

make another menu selection.

MOLYBDENUM, High Range

SMART3 Test Procedures 11.10

Test P

rocedures