Procedure – LaMotte SMART3 Colorimeter User Manual

Page 175

background image

PROCEDURE

1. Press and hold

until colorimeter turns on.

2. Press

ENTER

to select TESTING MENU.

3. Select

ALL TESTS (or another sequence containing 038 Dissolved Oxygen)

from TESTING MENU.

4. Scroll to and select 038 Dissolved Oxygen from menu.
5. Rinse a clean tube (0290) with untreated sample water. Fill to the 10 mL line

with sample. This tube is the BLANK.

6. Insert tube into chamber, close lid and select SCAN BLANK.
7. Fill a second tube (0290) to the 10 line with the treated “Fixed” sample. This

tube is the SAMPLE.

8. Remove BLANK from colorimeter, insert SAMPLE tube into chamber, close lid

and select SCAN SAMPLE. Record result.

9. Press

to turn colorimeter off or press

EXIT

to exit to a previous menu or

make another menu selection.

SMART3 Test Procedures 11.10

DISSOLVED OXYGEN