beautypg.com

Procedure – LaMotte SMART3 Colorimeter User Manual

Page 179

background image

PROCEDURE

1. Press and hold

until colorimeter turns on.

2. Press

ENTER

to select TESTING MENU.

3. Select

ALL TESTS (or another sequence containing 040 Fluoride) from

TESTING MENU.

4. Scroll to and select 040 Fluoride from menu.
5. This test requires a reagent blank. Rinse a clean tube (0290) with clear,

colorless, fl uoride free water. Fill to the 10 mL line with clear, colorless, fl uoride

free water.

6. Use the 0.5 mL pipet (0353) to add 0.5 mL of *Sodium Arsenite Solution

(4128). Cap and mix.

7. Use the 1.0 mL pipet (0354) to add 2 measures of *Acid-Zirconyl SPADNS

Reagent (3875). Cap and mix thoroughly. (This is the reagent blank.)

8. Insert tube into chamber, close lid and select SCAN BLANK.
9. Rinse a clean tube (0290) with sample water. Fill to the 10 mL line with sample

water. Repeat steps 6 and 7 .

10. Insert tube into chamber, close lid and select SCAN SAMPLE. Record result.
11. Press

to turn colorimeter off or press

EXIT

to exit to a previous menu or

make another menu selection.

SMART3 Test Procedures 11.13

FLUORIDE

Test P

rocedures