Procedure – LaMotte SMART3 Colorimeter User Manual

Page 200

background image

PROCEDURE

1. Press and hold

until colorimeter turns on.

2. Press

ENTER

to select TESTING MENU.

3. Select

ALL TESTS (or another sequence containing 053 Iron Phenthro) from

TESTING MENU.

4. Scroll to and select 053 Iron Phenthro from menu.
5. Rinse a clean tube (0290)with sample water. Fill to the 10 mL mark with

sample.

6. Insert tube into chamber, close lid and select SCAN BLANK.
7. Remove the tube from colorimeter. Remove the cap and add 6 drops of

*Acid Phenanthroline Indicator (2776). Cap and invert the tube 4 times to mix

reagents. Wait fi ve minutes for maximum color development.

8. After

fi ve minutes, mix, insert tube into chamber, close lid and select SCAN

SAMPLE. Record result as ppm Ferrous Iron.

9. Remove the tube from colorimeter. Use the 0.1g spoon (0699) to add one

measure of *Iron Reducing Reagent (2777). Cap and invert 15 times times to

mix. Wait 5 minutes for maximum color development.

10. After 5 minutes, mix, insert tube into chamber, close lid and select SCAN

SAMPLE. Record result as ppm Total Iron.

11. Press

to turn colorimeter off or press

EXIT

to exit to a previous menu or

make another menu selection.

12. Total Iron (ppm) - Ferrous Iron (ppm) = Ferric Iron (ppm)
NOTE: For best possible results, a reagent blank should be determined to account

for any contribution to the test result by the reagent system. To determine the

reagent blank, follow the above test procedure to scan a distilled or deionized

water blank. Then follow the above procedure to perform the test on a distilled or

deionized water sample. This test result is the reagent blank. Subtract the reagent

blank from all subsequent test results of unknown samples. It is necessary to

determine the reagent blank only when a new lot number of reagents is obtained.

IRON, Total, Ferrous, Ferric

SMART3 Test Procedures 11.10