PREPARING DILUTE STANDARD SOLUTIONS

Standard solutions should be prepared to create a calibration curve. Standard

solutions can be prepared by diluting a known concentrated standard by

specifi ed amounts. A chart or computer spreadsheet can be created to

determine the proper dilutions. Use volumetric fl asks and volumetric pipets for

all dilutions.

1. In Column A – Record the maximum concentration of test as determined by

the range and path length.

2. In Column B – Record the percent of the maximum concentration the

standard solution will be.

3. In Column C – Calculate the fi nal concentration of the diluted standard

solutions by multiplying the maximum concentration (In Column A) by the %

of maximum concentration divided by 100. (C = A x

B

/

100

).

4. In Column D – Record the fi nal volume of the diluted sample (i.e. volume of

volumetric fl ask).

5. In Column E – Record the concentration of the original standard.

6. In Column F – Calculate the milliliters of original standard required (F = (C x

D

/

E

)).

A sample chart appears below:

A

B

C =

A x

B

/

100

D

E

F =

C x

D

/

E

Maximum

concentration

of test

% of Maximum

concentration

Final

concentration

of Diluted

Standard

Volume of

Standard

Concentration

of Original

Standard

mL of

Original

Standard

Required

10.0 ppm

90

9.0 ppm

100 mL

1000 ppm

0.90 mL

10.0 ppm

70

7.0 ppm

100 mL

1000 ppm

0.70 mL

10.0 ppm

50

5.0 ppm

100 mL

1000 ppm

0.50 mL

10.0 ppm

30

3.0 ppm

100 mL

1000 ppm

0.30 mL

10.0 ppm

10

1.0 ppm

100 mL

1000 ppm

0.10 mL

10.0 ppm

0

0 ppm

100 mL

1000 ppm

0 mL