Procedure – LaMotte SMART3 Colorimeter User Manual

Page 263

background image

PROCEDURE

Use COD/UDV adapter.

1. Preheat COD reactor to 150 ±2°C. Follow safety precautions.
2. Remove cap from a *Total Phosphorus Acid Reagent Tube (4035). Use a 1.0

mL pipet (0354) to add 5.0 mL of sample.

3. Use the 0.15 g spoon (0727) and a funnel (0459) to add one level measure of

*Digestion Reagent Powder (4036). Tap funnel to dispense powder completely.

Cap tube tightly and shake until powder completely dissolves.

4. Place the tube in the COD reactor for 30 minutes.
5. At the end of the heating period, turn the reactor off. Carefully remove the tube

from the reactor and allow it to cool to room temperature.

6. At the end of the cooling period, press and hold

until colorimeter turns

on.

7. Press

ENTER

to select TESTING MENU.

8. Select

ALL TESTS (or another sequence containing 081 Phosphate T LR)

from TESTING MENU.

9. Scroll to and select 081 Phosphate T LR from the menu.
10. Carefully remove the cap from the digested tube. Use another 1 mL pipet

(0354) to add 1.0 mL of *Total Phosphorus LR Hydroxide Reagent (4038) to the

tube. Cap and invert to mix.

11. Wipe the tube with a damp towel to remove fi ngerprints and smudges. Wipe

with a dry towel.

12. Insert the tube into the chamber. Select SCAN BLANK. Remove the tube from

the colorimeter.

13. Use another 1 mL pipet (0354) to add *1.0 mL of Phosphate Acid Reagent (V-

6282). Cap and invert tube to mix.

14. Use the 0.1g spoon (0699) and a funnel (0459) to add one level spoon of

Phosphate Reducing Reagent (V-6283). Tap funnel to dispense powder

completely. Cap tube and shake until powder dissolves.

15. Wait 5 minutes.
16. Wipe the vials with a damp towel to remove fi ngerprints and smudges. Wipe

with a dry towel.

17. Insert the tube into the chamber. Select SCAN SAMPLE. Record the result as

Total Phosphorus in mg/L PO

4

.

18. Press

to turn the colorimeter off or press

EXIT

to exit to a previous

menu or make another menu selection.

NOTES: For greater accuracy, use laboratory grade pipets.

SMART3 Test Procedures 11.10

PHOSPHORUS, Total, Low Range

Test P