LaMotte SMART3 Colorimeter User Manual

Page 36

background image

5. Press

or

to scroll

to desired wavelength.

Absorbance

101 Absorbance 428

102 Absorbance 525

103 Absorbance 568
104 Absorbance 635

12:00:00 001/500

6. Press

ENTER

to select the

wavelength.

102 Absorbance 525

Scan Blank

Scan Sample

12:00:00 001/500

7. Insert the blank. Close the lid.

Press

ENTER

to scan the blank.

The screen wil display

Blank

Done for about 1 second

and return to the Absorbance

menu.

102 Absorbance 525

Scan Blank

Scan Sample

12:00:00 001/500

8. Insert the reacted sample.

Press

ENTER

to scan the

sample. The screen will

display

Reading for about 1

second. The result will appear

on the screen.

102 Absorbance 525

0.425

Scan Blank

Scan Sample

12:00:00 001/500

34

SMART3 Colorimeter 11.10

Testing