Servicing the at30 – Exide Technologies Section 94.40 User Manual
Page 52

SERVICING THE AT30
48
3.3. USING THE TROUBLESHOOTING CHART
WARNING 
High voltages appear at several points inside the AT30. Use extreme 
caution when working inside the unit. Do not attempt to work inside the 
AT30 unless you are a qualified technician or electrician.  
Disconnect and lock out all power from the AT30 before starting to 
remove or replace any components. Turn the ac power off at the 
distribution panel upstream from the charger. Disconnect the battery 
from the AT30 output terminals TB1(+/-). 
Before you try to use the troubleshooting chart, be sure that you have 
followed the steps in Section 3.1. 
The troubleshooting chart that begins on the next page is divided into 
three columns. To use the chart: 
1. Make a list of the charger's condition, including the trouble symptoms.
2. Find the symptom(s) in the first column of the chart.
3. The middle column contains common causes for the problem you observe, in
order of probability.
4. Follow the action described in the right-hand column to correct the problem
and return the charger to normal service.
Determining the condition of the charger:
•
Is the front panel
AC ON indicator lit?
•
What is the ac voltage at the input terminals (measured with an ac 
voltmeter)? 
•
What is the dc voltage at the output terminals (measured with a dc 
voltmeter)? 
•
Does the meter on the front panel display any voltage or current?
•
Are any alarm indicators lit?
•
Do the front panel status indicators work (Charge Mode, for example)?
•
Can you change the charger to the Equalize mode and back to Float?
•
Is the charger making any noise? Is it unusually loud?
•
Is there any sign or smell of smoking or burning?
For instructions on replacing components, see Section 3.5 on page 58. For 
information on identifying internal components, refer to the standard 
drawings featured in Appendix C. 
