7 How to ...?
VA Database 2.0
71
Note: Each time a determination is sent to the database, a de-
termination file (
*.dtm
) and an ASCII determination report file (
*.rpt
)
will be stored automatically to the backup directory if the option
Save original files
is enabled in the
SERIAL INTERFACE SETTINGS
window (see Default directories, page 51).
Send methods to the backup directory (binary and
ASCII files)
1. If you want to observe the process of data transfer, click on
or
MAIN WINDOW / Window / Communication
to open the
VA
COMMUNICATION
window (the data transfer works also in the
background).
2. Load the method to be sent into the working storage of the
746 VA Trace Analyzer.
3. On dialog page "DOCUMENTATION" in the 746 VA Trace
Analyzer enter on a new line
Report ActMeth
in the
COPY Re-
ports, Curves
column and
RSIfc.1
or
RSIfc.2
in the
TO destina-
tion
column.
4. Activate the data transfer by moving the cursor to the de-
scribed line and pressing the softkey
.
Using this procedure both the binary method file (
*.mth
) and the
ASCII method report file (
*.rpt
) are transferred and stored in the
backup directory.
Note: If a method contains the line
COPY
Report ActMeth TO
RSIfc.#
on the dialog page "DOCUMENTATION", the method files
are sent automatically to the backup directory after termination of
the analysis.
Send methods to the backup directory (only binary
files)
1. If you want to observe the process of data transfer, click on
or
MAIN WINDOW / Window / Communication
to open the
VA
COMMUNICATION
window (the data transfer works also in the
background).
2. Select the method(s) to be sent on the dialog page
"METHODS" or "DATA CARD" in the 746 VA Trace Analyzer.
3. Press the softkeys
and
/
.
The binary method files (
*.mth
) will be transferred to the prede-
fined backup directory (see Default directories, page 51).