8 Troubleshooting
VA Database 2.0
94
The automatic template load cannot be activated
1. Press
to change to the read-only mode in the
VA REPORT
GENERATOR
window.
2. Save the current template which has been modified.
The custom reports are not printed with the currently
loaded template
Deactivate the automatic loading of method associated tem-
plates.
The automatic custom report printing is not working
1. Check if automatic printing is enabled in the
SERIAL INTER-
FACE SETTINGS
window (see Automatic printing, page 53).
2. Check printing condition in the
SERIAL INTERFACE SETTINGS
window (see Automatic printing, page 53).
Not all custom reports are printed with manual
selection and automatic template load enabled
1. Check if for all methods used in the determinations to be
printed a method associated template is defined in the
GEN-
ERAL PREFERENCES
window (see Templates settings, page
22).
2. Define a method associated template for each method used
or disable the automatic loading of templates.
8.3 Display: What to do if ...
No overlay curves are created
Check if the selected items are compatible. Overlay curves
can only be created if they are of the same type and have the
same units for x and y axes.
Examples for impossible overlays are:
substance and segment curves,
standard addition and calibration curves,
calibration curves with height and area evaluation,
VA and PSA curves.