8 Troubleshooting

VA Database 2.0

93

2. Check the settings for text report printing in the

PAGE SETUP

window (see Document page setup, page 11). Select the tab

Documentation

and set the

Report type

to

All

if you want to print

the whole text report.

Curves are missing in curve printing

1. Check the color settings for curve lines (some colors may

vanish on a black and white printer). Select black or dark col-
ors as default in the

GENERAL PREFERENCES

window (see

Settings, page 19) or

OPTIONS

window (see Display of curves,

page 14).

2. Check the curve subwindow in the

MAIN WINDOW

. If no curve

is visible, the scaling may be wrong. Adjust the scaling in the

OPTIONS

window (see Scaling of curve axes, page 18) or re-

set the curve window to the default scale by pressing the
key or select the

Set Default Scale

option of the context

sensitive menu.

3. If the curve is not visible with default scale settings, change

the x and y axes scales to

Auto

in the

OPTIONS

window (see

Display of curves, page 14).

4. Since the changed scaling settings are not stored in the da-

tabase, you should correct the wrong scaling settings in the

746 VA Trace Analyzer. For this, send the selected determi-

nation to the 746 VA Trace Analyzer and check the

Dis-

play/Plot

parameters for segments or substances. Correct the

scaling to the desired values and send the determination

back to the VA Database.

Curves in custom reports are missing in printing

1. Check the color settings for curve lines (some colors may

vanish on a black and white printer). Select black or dark col-
ors as default in

OPTIONS

window (see Curve fields, page 64).

2. If the curve is still not visible, send the selected determination

to the 746 VA Trace Analyzer and check the

Display/Plot

pa-

rameters for segments or substances. Correct the scaling to
the desired values and send the determination back to the VA
Database.

Curve and axes lines are very thin

This problem may occur with printers with a high graphics
resolution. Reduce the graphics resolution on your printer
(e.g. to 600 dpi or 300 dpi).