Metrohm NIRS XDS RapidLiquid Analyzer User Manual

Page 55

background image

▪▪▪▪▪▪▪

53

Vision requests a tolerance file. The
tolerance file was loaded in the C:\Vision
directory, and is an “XDA” formatted file.

Select this file and click “Open”.

Current versions of Vision include a newer
version of this tolerance file, which may be
named “IPVXDS02, 03, or higher. These files
support additional XDS instruments, though
that will not be visible to the user. The
tolerances for the XDS Rapid Liquid Analyzer
remain the same in all versions.

Vision displays the wavelength regions for
test.

For Rapid Liquid instruments, these regions
give a good overall picture of instrument
performance and repeatability. The
wavelength areas are chosen in flat parts of
the standards spectra for stability.

For Number of Replicates, retain the default
setting of 1.

Click on “OK”.

Vision will take an instrument reference
scan, with the sample drawer in the open
position. The red progress bar at the bottom
of the screen indicates status.

Vision requests the T01 standard from the
set, as shown.