Metrohm 761 SD Compact IC User Manual

Page 90

background image

5 Operation

761 SD Compact IC / Instructions for Use 8.

761.1043

82

5.1.2.2

Calibration and sample determination (manual operation / «IC
Cap»)

The instrument should be calibrated after each restart.

The sample aspirating tubing

3

should be immersed in the PE bottle

(6.1608.080) filled with ultra-pure water after each sample (or stan-

dard).

1 Set system manual.smt

Click on the yellow <Sample> button at the top right so that

it changes to the green <Start> button.

Choose system

manual.smt

for item

System

.

2 Calibration

Insert the free (dry) end of the sample aspirating tubing 3 into

the MCAL standard with the lowest concentration (150 mg/L

H

3

PO

4

plus basic content of nitrate and sulfate). The MCAL

standard bottles are supplied as accessories (6.2321.000).

Choose type "Calib(1)" for item

Run Type

.

Enter a selection, e.g. "Standard 1" for item

Ident

.

Start the determination by clicking on the green but-

ton and allow it to run through to the end.

After completion of this measurement, you should determine
the other 5 MCAL standards one after the other (250, 350,
450, 550 and 650 mg/L H

3

PO

4

plus basic content of nitrate

and sulfate). When doing this, increase the type in increments
in each case by 1 (Calib(X)) in item

Run Type

and adjust the

entry in item

Ident

(this means that type "Calib(6)" should be

set for the

Run Type

and “Standard 6” should be set for item

Ident

in the above example for measurement with the 650

mg/L standard).

In order to check calibration, you should also determine the

MCAL QC check standard with 545 mg/L H

3

PO

4

plus basic

content of nitrate and sulfate

at the end of the series of cali-

bration measurements. Choose type “Sample” in item

Run

Type

for this measurement.