5 Operation
761 SD Compact IC / Instructions for Use 8.
761.1043
80
3 Sample determination
⇒
System
manual.smt
(see Section 3.2.2) should still be open
and linked from calibration. If not, open it (see Section 3.1.2)
and link it (see Section 3.1.4).
⇒
Dry off the free end of the sample aspirating tubing 3 with a
clean cloth.
⇒
Then insert it into the sample vessel.
⇒
Start the system by selecting
SYSTEM /
Control
/ Start determi-
nation
(see Software Instructions for Use «IC Net 2.3», Sections
4.3.3 and 4.3.4). Set the
Calibration level
in window
EDIT
SAMPLE DESCRIPTION: SYSTEMS
to 0.
⇒
Information on chromatograms and post-processing chroma-
tograms can be found in the Software Instructions for Use «IC
Net 2.3», Section 8 resp. Section 10.
5.1.1.3
Shutdown (manual operation / «IC Net»)
In order to prevent damage to the instrument, it should be rinsed with
water before shutting it down after the end of sample determination.
This rinsing operation should be performed with system
shutdown.smt
(see Section 3.2.4).
1 Switch to system shutdown.smt and link it
⇒
Switch (see Software Instructions for Use «IC Net 2.3», Section
4.2) to system
shutdown.smt
(see Section 3.2.2).