5 Operation

761 SD Compact IC / Instructions for Use 8.

761.1043

80

3 Sample determination

System

manual.smt

(see Section 3.2.2) should still be open

and linked from calibration. If not, open it (see Section 3.1.2)

and link it (see Section 3.1.4).

Dry off the free end of the sample aspirating tubing 3 with a

clean cloth.

Then insert it into the sample vessel.

Start the system by selecting

SYSTEM /

Control

/ Start determi-

nation

(see Software Instructions for Use «IC Net 2.3», Sections

4.3.3 and 4.3.4). Set the

Calibration level

in window

EDIT

SAMPLE DESCRIPTION: SYSTEMS

to 0.

Information on chromatograms and post-processing chroma-

tograms can be found in the Software Instructions for Use «IC

Net 2.3», Section 8 resp. Section 10.

5.1.1.3

Shutdown (manual operation / «IC Net»)

In order to prevent damage to the instrument, it should be rinsed with

water before shutting it down after the end of sample determination.

This rinsing operation should be performed with system

shutdown.smt

(see Section 3.2.4).

1 Switch to system shutdown.smt and link it

Switch (see Software Instructions for Use «IC Net 2.3», Section

4.2) to system

shutdown.smt

(see Section 3.2.2).