3.1 «IC Net» – User interface for the 761 SD Compact IC

761 SD Compact IC / Instructions for Use 8.

761.1043

47

Peristaltic pump

Peristaltic

pump

Switching on the pump drive.

Switching off the pump drive.

System startup values

System startup values. These parameters are

set on the 761 SD Compact IC when linking the

system, when starting a determination process

or when sending manually with

.

Flow, mL/min

Startup value for flow rate of the high-pressure

pump.

Entry range:

0.20

...

2.50 mL/min

Pressure max, MPa

Startup value for maximum switch-off pressure

for the high-pressure pump. (This value is

monitored once it has been set even without a

connection to the PC).

Entry range:

0.0

...

25.0 MPa

Pressure min, MPa

Startup value for minimum switch-off pressure

for the high-pressure pump. This value is moni-

tored even without a connection to the PC.

Full scale,

µ

S/cm

Startup value for full-scale range.

Selection:

50

,

250

,

1000

µ

S/cm

Remote lines

Startup values for status of the remote output

lines 1

8.

Selection:

0

,

1

Time program

A user-specific time program for instrument control can be entered on

the

Program

page in the window for the system settings. This program is

automatically started either when starting the determination process

(

Start with determination

) or on injection of the sample (

Start with inject

),

depending on the setting in window

Start mode

(see Software Instruc-

tions for Use «IC Net 2.3», Section 4.3.3).