Metrohm 761 SD Compact IC User Manual
Page 55
3.1 «IC Net» – User interface for the 761 SD Compact IC
761 SD Compact IC / Instructions for Use 8.
761.1043
47
Peristaltic pump
Peristaltic
pump
Switching on the pump drive.
Switching off the pump drive.
System startup values
System startup values. These parameters are
set on the 761 SD Compact IC when linking the
system, when starting a determination process
or when sending manually with
.
Flow, mL/min
Startup value for flow rate of the high-pressure
pump.
Entry range:
0.20
...
2.50 mL/min
Pressure max, MPa
Startup value for maximum switch-off pressure
for the high-pressure pump. (This value is
monitored once it has been set even without a
connection to the PC).
Entry range:
0.0
...
25.0 MPa
Pressure min, MPa
Startup value for minimum switch-off pressure
for the high-pressure pump. This value is moni-
tored even without a connection to the PC.
Full scale,
µ
S/cm
Startup value for full-scale range.
Selection:
50
,
250
,
1000
µ
S/cm
Remote lines
Startup values for status of the remote output
lines 1
…
8.
Selection:
0
,
1
Time program
A user-specific time program for instrument control can be entered on
the
Program
page in the window for the system settings. This program is
automatically started either when starting the determination process
(
Start with determination
) or on injection of the sample (
Start with inject
),
depending on the setting in window
Start mode
(see Software Instruc-
tions for Use «IC Net 2.3», Section 4.3.3).