5.1 Operation with manual sample change
761 SD Compact IC / Instructions for Use 8.
761.1043
79
2 Calibration
⇒
Dry off the free end of the sample aspirating tubing 3 with a
clean cloth.
⇒
Then immerse it in the lowest MCAL standard (mg/L H
3
PO
4
plus basic content of nitrate and sulfate). The MCAL standard
bottles are supplied as accessories (6.2321.000).
⇒
Start the system by selecting
SYSTEM /
Control
/ Start determi-
nation
(see Software Instructions for Use «IC Net 2.3», Sections
4.3.3 and 4.3.4). After start, you will see the
EDIT SAMPLE
DESCRIPTION: SYSTEMS
window. You should make the fol-
lowing entries:
What is important above all is to set menu option
Calibration
level
to 1.
⇒
After completing this measurement, you should determine the
other 5 MCAL standards one after the other (250, 350, 450,
550 and 650 mg/L H
3
PO
4
plus basic content of nitrate and
sulfate) and, when doing this, increase menu item
Calibration
level
in window
EDIT SAMPLE DESCRIPTION: SYSTEMS
in-
crementally by one (this means that the measurement with
the 650 mg/L standard would then be
Calibration level
6).
⇒
In order to check calibration, you should also determine the
MCAL QC check standard with 545 mg/L H
3
PO
4
plus basic
content of nitrate and sulfate at the end of the series of cali-
bration measurements.
⇒
Detailed information on calibration can be found in the Soft-
ware Instructions for Use «IC Net 2.3», Section 7.5.