2 Installation

761 SD Compact IC / Instructions for Use 8.

761.1043

36

1 Detach the column connection capillary 33 from the

PEEK coupling

Undo the pressure screw 50 with which the column connec-

tion capillary 33 is screwed to the PEEK coupling 54 (transi-

tion to the suppressor). Take this end of the capillary and

route it into a beaker.

The column connection capillary 33 should now connect the in-

jection valve to the beaker.

2 Open and connect the system

Open the

startup.smt

system, connect it and make the same

settings as described in Section 2.13.1.

3 Rinse the pulsation dampener

Click on the

button for

IC pump

to switch on the high-

pressure pump and rinse the pulsation dampener 30 filled

with isopropanol for approx. 10 minutes with eluent.

Click on the

button for

IC pump

to switch the high-

pressure pump back off again.

4 Connect the column connection capillary 33 back to

the PEEK coupling

Take the end of the column connection capillary 33 out of the

beaker and screw it back onto the PEEK coupling 54 (as a

transition to the suppressor) with a pressure screw 50.